このエントリーをはてなブックマークに追加
ID 13843
Eprint ID
13843
FullText URL
Author
Noguchi, Osamu
Oshima, Mitsuko
Motomizu, Shoji
Abstract
The application of inductively coupled plasma mass spectrometry (ICP-MS) to forensic chemistry was studied. The developed method, air-segmented sample injection (ASSI) coupled with ICP-MS, allowed the determination of about 25 elements at the sub-ppb level with only 0.2 ml of a sample solution. The optimum sample flow rate was found to be 0.4 ml min–1, along with a sample suction time of 30 s. The proposed method was validated by determining trace elements in river-water certified reference material (SLRS-4) issued by National Research Council Canada. The analytical results of the proposed method were in good agreement with the certified values. This method was successfully applied to a human hair sample, the volume of which was 3 ml.
Published Date
2008-05-10
Publication Title
Analytical Sciences
Volume
volume24
Issue
issue5
Publisher
日本分析化学会
Publisher Alternative
Japan Society for Analytical Chemistry
Start Page
631
End Page
635
ISSN
0910-6340
NCID
AA10500785
Content Type
Journal Article
language
English
Copyright Holders
日本分析化学会
File Version
publisher
Refereed
True
DOI