ID | 13843 |
Eprint ID | 13843
|
フルテキストURL | |
著者 |
Noguchi, Osamu
大島 光子
岡山大学
本水 昌二
岡山大学
|
抄録 | The application of inductively coupled plasma mass spectrometry (ICP-MS) to forensic chemistry was studied. The
developed method, air-segmented sample injection (ASSI) coupled with ICP-MS, allowed the determination of about 25
elements at the sub-ppb level with only 0.2 ml of a sample solution. The optimum sample flow rate was found to be 0.4
ml min–1, along with a sample suction time of 30 s. The proposed method was validated by determining trace elements in river-water certified reference material (SLRS-4) issued by National Research Council Canada. The analytical results of the proposed method were in good agreement with the certified values. This method was successfully applied to a human hair sample, the volume of which was 3 ml.
|
発行日 | 2008-05-10
|
出版物タイトル |
Analytical Sciences
|
巻 | 24巻
|
号 | 5号
|
出版者 | 日本分析化学会
|
出版者(別表記) | Japan Society for Analytical Chemistry
|
開始ページ | 631
|
終了ページ | 635
|
ISSN | 0910-6340
|
NCID | AA10500785
|
資料タイプ |
学術雑誌論文
|
言語 |
英語
|
著作権者 | 日本分析化学会
|
論文のバージョン | publisher
|
査読 |
有り
|
DOI |