このエントリーをはてなブックマークに追加


ID 30187
FullText URL
Author
Nakata, Takayoshi
Takahashi, N.
Fujiwara, K.
Nakano, Masanori
Ogura, Y.
Abstract

When the DC magnetization curve (B-H) of nonoriented material is measured in a ring specimen, there is an intrinsic error due to the assumption that the mean magnetic path length is equal to the mean geometric path length. A novel method for determining the B-H curve accurately is proposed. The validity of the method is verified by experiments

Keywords
magnetic variables measurement
magnetisation
Note
Digital Object Identifier: 10.1109/20.179524
Published with permission from the copyright holder. This is the institute's copy, as published in Magnetics, IEEE Transactions on, Sep. 1992, Volume 28, Issue 5, Part 2, Pages 2456-2458.
Publisher URL:http://dx.doi.org/10.1109/20.179524
Copyright © 1992 IEEE. All rights reserved.
Published Date
1992-9
Publication Title
Magnetics
Volume
volume28
Issue
issue5
Start Page
2456
End Page
2458
Content Type
Journal Article
language
English
Refereed
True
DOI
Submission Path
electricity_and_magnetism/66