Abstract
We developed microfluidic paper-based analytical devices (μPADs) for the chelate titrations of Ca2+ and Mg2+ in natural water. The μPAD consisted of ten reaction zones and ten detection zones connected through narrow channels to a sample zone located at the center. Buffer solutions with a pH of 10 or 13 were applied to all surfaces of the channels and zones. Different amounts of ethylenediaminetetraacetic acid (EDTA) were added to the reaction zones and a consistent amount of a metal indicator (Eriochrome Black T or Calcon) was added to the detection zones. The total concentrations of Ca2+ and Mg2+ (total hardness) in the water were measured using a μPAD containing a buffer solution with a pH of 10, whereas only Ca2+ was titrated using a μPAD prepared with a potassium hydroxide solution with a pH of 13. The μPADs permitted the determination of Ca2+ and Mg2+ in mineral water, river water, and seawater samples within only a few minutes using only the naked eye—no need of instruments.
Keywords
Chelate titration
Microfluidic paper-based analytical device
Hardness
Calcium
Magnesium
Natural water
Note
© 2016. This manuscript version is made available under the CC-BY-NC-ND 4.0 license http://creativecommons.org/licenses/by-nc-nd/4.0/
Published Date
2016-06-14
Publication Title
Analytica Chimica Acta
Volume
volume924
Publisher
Elsevier B.V.
Start Page
60
End Page
67
ISSN
0003-2670
NCID
AA00524834
Content Type
Journal Article
language
English
OAI-PMH Set
岡山大学
Copyright Holders
© 2016 Elsevier B.V.
File Version
author
PubMed ID
DOI
Web of Science KeyUT
Related Url
isVersionOf https://doi.org/10.1016/j.aca.2016.04.019