このエントリーをはてなブックマークに追加


ID 32186
JaLCDOI
FullText URL
Author
Furuno, Katsushi
Gomita, Yutaka
Araki, Yasunori
Fukuda, Tamotu
Abstract

We studied the use of high-performance liquid chromatography (HPLC), using a solid phase extraction column (Bond Elut cartridge column), for the simple, rapid and sensitive determination of serum clonazepam levels in epileptic patients. Extracted aliquots were analyzed by HPLC, using a reverse phase ODS column (mu-Bondapak C18). The analytical mean recovery of clonazepam added to the blank serum averaged 99.9%. The detection limit was as high as approximately 2 ng/ml in the serum. The reproducibilities were 2.3-8.6 CV % in the within-day assay and 6.5 CV % in the between-day assay, indicating that the analysis method was effective in the determination of clonazepam serum levels. Accordingly, we suggest that the present method, using a solid phase extraction column, may be useful for the routine monitoring of clonazepam serum levels in epileptic patients.</P>

Keywords
clonazepam serum levels
epileptic patient
therapeutic drug monitoring
solid-phase extraction
HPLC
Amo Type
Article
Publication Title
Acta Medica Okayama
Published Date
1991-04
Volume
volume45
Issue
issue2
Publisher
Okayama University Medical School
Start Page
123
End Page
127
ISSN
0386-300X
NCID
AA00508441
Content Type
Journal Article
language
English
File Version
publisher
Refereed
True
PubMed ID
Web of Science KeyUT