Abstract | We studied the use of high-performance liquid chromatography (HPLC), using a solid phase extraction column (Bond Elut cartridge column), for the simple, rapid and sensitive determination of serum clonazepam levels in epileptic patients. Extracted aliquots were analyzed by HPLC, using a reverse phase ODS column (mu-Bondapak C18). The analytical mean recovery of clonazepam added to the blank serum averaged 99.9%. The detection limit was as high as approximately 2 ng/ml in the serum. The reproducibilities were 2.3-8.6 CV % in the within-day assay and 6.5 CV % in the between-day assay, indicating that the analysis method was effective in the determination of clonazepam serum levels. Accordingly, we suggest that the present method, using a solid phase extraction column, may be useful for the routine monitoring of clonazepam serum levels in epileptic patients.</P> |
Keywords | clonazepam serum levels
epileptic patient
therapeutic drug monitoring
solid-phase extraction
HPLC
|
Amo Type | Article
|
Publication Title |
Acta Medica Okayama
|
Published Date | 1991-04
|
Volume | volume45
|
Issue | issue2
|
Publisher | Okayama University Medical School
|
Start Page | 123
|
End Page | 127
|
ISSN | 0386-300X
|
NCID | AA00508441
|
Content Type |
Journal Article
|
language |
English
|
File Version | publisher
|
Refereed |
True
|
PubMed ID | |
Web of Science KeyUT |