Abstract
We developed an organic solvent-compatible paper-based analytical device (PAD) for the quantitative analysis of indole, which is an indicator of shrimp freshness. Although indole is insoluble in water, ethyl acetate is a suitable solvent to dissolve and extract indole from shrimp. The PADs are fabricated using a cutting method that allows the use of an organic solvent because no hydrophobic barrier is needed to form fluidic channels. Ehrlich's reagent consists of 4-(dimethylamino)benzaldehyde and p-dimethylaminobenzaldehyde and was deposited onto the reaction zone of the PAD followed by lamination to prevent evaporation of the ethyl acetate. Samples are introduced into the PAD via immersion in organic sample solutions. When the PAD is immersed into an indole solution of ethyl acetate in a closed bottle, the sample solution penetrates the channel of the PAD and successively flows into the detection zone to form a hydrophilic colored product. The PADs provide a linear relationship between the logarithm of the indole concentration and the color intensity within a range of 1.0-20 ppm with correlation coefficients of r2 > 0.99. The limits of detection and quantification are 0.36 and 0.71 ppm, respectively. Relative standard deviations for both the intraday (n = 2) and interday (n = 3) precision were less than 2.5%. In the indole analysis of shrimp, the PADs separated the interfering orange-colored astaxanthin in the extract from the colored product of indole via the paper chromatographic principle. We used the PADs to investigate the degradation of shrimp, and the results showed a rapid increase in the indole level after 7 days. High-performance liquid chromatography verified the accuracy of the PADs by showing good agreement with the obtained indole levels.
Keywords
Ehrlich’s reagent
astaxanthin
chromatography
indole
paper-based device
shrimp
Note
This document is the Accepted Manuscript version of a Published Work that appeared in final form in ACS Sensors, copyright © 2022 American Chemical Society after peer review and technical editing by the publisher. To access the final edited and published work see https://doi.org/10.1021/acssensors.2c00300
Published Date
2022-4-11
Publication Title
ACS Sensors
Volume
volume7
Issue
issue4
Publisher
American Chemical Society
Start Page
1194
End Page
1200
ISSN
2379-3694
Content Type
Journal Article
language
English
OAI-PMH Set
岡山大学
Copyright Holders
© 2022 American Chemical Society
File Version
author
PubMed ID
DOI
Web of Science KeyUT
Related Url
isVersionOf https://doi.org/10.1021/acssensors.2c00300
Citation
Dip-and-Read, Organic Solvent-Compatible, Paper-Based Analytical Devices Equipped with Chromatographic Separation for Indole Analysis in Shrimp Sasikarn Seetasang and Takashi Kaneta ACS Sensors 2022 7 (4), 1194-1200