ID | 63703 |
FullText URL | |
Author |
Seetasang, Sasikarn
Department of Chemistry, Graduate School of Natural Science and Technology, Okayama University
Kaneta, Takashi
Department of Chemistry, Graduate School of Natural Science and Technology, Okayama University
ORCID
Kaken ID
publons
researchmap
|
Abstract | We developed an organic solvent-compatible paper-based analytical device (PAD) for the quantitative analysis of indole, which is an indicator of shrimp freshness. Although indole is insoluble in water, ethyl acetate is a suitable solvent to dissolve and extract indole from shrimp. The PADs are fabricated using a cutting method that allows the use of an organic solvent because no hydrophobic barrier is needed to form fluidic channels. Ehrlich's reagent consists of 4-(dimethylamino)benzaldehyde and p-dimethylaminobenzaldehyde and was deposited onto the reaction zone of the PAD followed by lamination to prevent evaporation of the ethyl acetate. Samples are introduced into the PAD via immersion in organic sample solutions. When the PAD is immersed into an indole solution of ethyl acetate in a closed bottle, the sample solution penetrates the channel of the PAD and successively flows into the detection zone to form a hydrophilic colored product. The PADs provide a linear relationship between the logarithm of the indole concentration and the color intensity within a range of 1.0-20 ppm with correlation coefficients of r2 > 0.99. The limits of detection and quantification are 0.36 and 0.71 ppm, respectively. Relative standard deviations for both the intraday (n = 2) and interday (n = 3) precision were less than 2.5%. In the indole analysis of shrimp, the PADs separated the interfering orange-colored astaxanthin in the extract from the colored product of indole via the paper chromatographic principle. We used the PADs to investigate the degradation of shrimp, and the results showed a rapid increase in the indole level after 7 days. High-performance liquid chromatography verified the accuracy of the PADs by showing good agreement with the obtained indole levels.
|
Keywords | Ehrlich’s reagent
astaxanthin
chromatography
indole
paper-based device
shrimp
|
Note | This document is the Accepted Manuscript version of a Published Work that appeared in final form in ACS Sensors, copyright © 2022 American Chemical Society after peer review and technical editing by the publisher. To access the final edited and published work see https://doi.org/10.1021/acssensors.2c00300
|
Published Date | 2022-4-11
|
Publication Title |
ACS Sensors
|
Volume | volume7
|
Issue | issue4
|
Publisher | American Chemical Society
|
Start Page | 1194
|
End Page | 1200
|
ISSN | 2379-3694
|
Content Type |
Journal Article
|
language |
English
|
OAI-PMH Set |
岡山大学
|
Copyright Holders | © 2022 American Chemical Society
|
File Version | author
|
PubMed ID | |
DOI | |
Web of Science KeyUT | |
Related Url | isVersionOf https://doi.org/10.1021/acssensors.2c00300
|
Citation | Dip-and-Read, Organic Solvent-Compatible, Paper-Based Analytical Devices Equipped with Chromatographic Separation for Indole Analysis in Shrimp Sasikarn Seetasang and Takashi Kaneta ACS Sensors 2022 7 (4), 1194-1200
|