Korenaga Takashi
Moriwake Tosio
Takahashi Teruo
Abstract
A simple system was developed for the fully automatic and continuous measurement of chemical oxygen demand (COD) in wastewater samples based on colorimetry of dichromate. A sample and a solution of sulfuric acid (1+1) containing 2mM potassium dichromate are continuously pumped with a double-reciprocating micro-pump at each flow rate of 0.3 ml/min. The wastewater sample is filtered at first with a 100-mesh stainless filter and then mixed with the dichromate solution in the mixing joint. The mixture is introduced into a reaction coil made of poly(tetrafluoroethylene) tubing (1 mm i.d., 3 mm o.d., and 20 m length), being placed in an oil bath (120℃). After reaction, the mixture passes into a quartz tubular flow-through cell (10 mm path length, 18 μl volume) in a spectrophotometer, and the absorbance is measured at 445 nm. The COD value of the sample is automatically estimated from the amount of decreased absorbance. The system was successfully applied to COD measurement of some waters, and to continuous monitoring of COD in wastewater of university laboratories. The system was also evaluated by comparing with the flow injection analyzer system previously developed by the authors.
Publication Title
Memoirs of the School of Engineering, Okayama University
Published Date
1985-11-08
Volume
volume20
Issue
issue1
Publisher
岡山大学工学部
Publisher Alternative
School of Engineering, Okayama University
Start Page
45
End Page
52
ISSN
0475-0071
NCID
AA00733903
Content Type
Departmental Bulletin Paper
OAI-PMH Set
岡山大学
language
English
File Version
publisher
NAID
Eprints Journal Name
mfe