Development of Computer-Controlled Flow Injection Instruments and Its Application to Determination of Nitrate, Nitrite, and Ammonium Ions in Environmental Samples
Author
Joichi, Yasutaka
Narong, Lenghor
Oshima, Mitsuko
Motomizu, Shoji
Ura, Nobuo
Abstract
Flow injection analysis (FIA) systems with low power consumption and personal-computer control were developed by using solenoid pumps and an LED detector. Both programs for controlling the system and acquiring measurement data were written using LabVIEW. The FIA system was applied to the determination of nitrite ion in river-water samples based on diazotization and coupling reactions. Linear calibration graphs were obtained with the combination of pump(s) and photometric detector in the concentration range of nitrite ion at 0.5∼3.0 × 10−6 M. Nitrate and nitrite ions were also determined simultaneously by the system by using a Cd/Cu reduction column and a by-pass flow. Determination of ammonia in river-water samples was also realized by the system with a modified indophenol reaction.
Keywords
Flow injection analysis
LabVIEW control
solenoid pumps
nitrate ion
nitrite ion
ammonia
Published Date
2006-11-17
Publication Title
分析化学
Publication Title Alternative
Bunseki Kagaku
Volume
volume55
Issue
issue9
Publisher
日本分析化学会
Publisher Alternative
The Japan Society for Analytical Chemistry
Start Page
707
End Page
713
ISSN
0525-1931
NCID
AN00222633
Content Type
Journal Article
language
Japanese
Copyright Holders
The Japan Society for Analytical Chemistry
File Version
publisher
Refereed
True
DOI