抄録
The aim of this study is to evaluate the patient positioning uncertainty in noncoplanar stereotactic radiosurgery or stereotactic radiotherapy (SRS/SRT) for intracranial lesions with the frameless 6D ExacTrac system. In all, 28 patients treated with SRS/SRT of 70 treatment plans at our institution were evaluated in this study. Two X-ray images with the frameless 6D ExacTrac system were first acquired to correct (XC) and verify (XV) the patient position at a couch angle of 0o. Subsequently, the XC and XV images were also acquired at each planned couch angle for using noncoplanar beams to detect position errors caused by rotating a couch. The translational XC and XV shift values at each couch angle were calculated for each plan. The percentages of the translational XC shift values within 1.0 mm for each planned couch angle for using noncoplanar beams were 77.86%, 72.26%, and 98.47% for the lateral, longitudinal, and vertical directions, respectively. Those within 2.0 mm were 98.22%, 97.96%, and 99.75% for the lateral, longitudinal, and vertical directions, respectively. The maximum absolute values of the translational XC shifts among all planned couch angles for using noncoplanar beams were 2.69, 2.45, and 2.17 mm for the lateral, longitudinal, and vertical directions, respectively. The overall absolute values of the translational XV shifts were less than 1.0 mm for all directions except for one case in the longitudinal direction. The patient position errors were detected after couch rotation for using noncoplanar beams, and they exceeded a planning target volume (PTV) margin of 1.0-2.0 mm used commonly in SRS/SRT treatment. These errors need to be corrected at each planned couch angle, or the PTV margin should be enlarged.
キーワード
IGRT
noncoplanar radiotherapy
patient positioning uncertainty
SRS
SRT
発行日
2020-01-20
出版物タイトル
Journal of Applied Clinical Medical Physics
21巻
2号
出版者
Wiley
開始ページ
89
終了ページ
97
ISSN
1526-9914
NCID
AA12034060
資料タイプ
学術雑誌論文
言語
英語
OAI-PMH Set
岡山大学
著作権者
© 2020 The Authors.
論文のバージョン
publisher
PubMed ID
DOI
Web of Science KeyUT
関連URL
isVersionOf https://doi.org/10.1002/acm2.12820
ライセンス
https://creativecommons.org/licenses/by/4.0/