ID | 67173 |
著者 |
Ando, Yasuharu
Hiroshima City Hospital
Okada, Masahiro
Hiroshima City North Medical Center Asa Citizens Hospital
Matsumoto, Natsuko
Hiroshima City North Medical Center Asa Citizens Hospital
Ikuhiro, Kawasaki
Hiroshima City North Medical Center Asa Citizens Hospital
Ishihara, Soichiro
Hiroshima City Hospital
Kiriu, Hiroshi
Hiroshima City Hospital
Tanabe, Yoshinori
Department of Radiological Technology, Graduate School of Health Sciences, Okayama University
ORCID
Kaken ID
researchmap
|
抄録 | This study aims to evaluate the output factors (OPF) of different radiation therapy planning systems (TPSs) using a plastic scintillator detector (PSD). The validation results for determining a practical field size for clinical use were verified. The implemented validation system was an Exradin W2 PSD. The focus was to validate the OPFs of the small irradiation fields of two modeled radiation TPSs using RayStation version 10.0.1 and Monaco version 5.51.10. The linear accelerator used for irradiation was a TrueBeam with three energies: 4, 6, and 10 MV. RayStation calculations showed that when the irradiation field size was reduced from 10 × 10 to 0.5 × 0.5 cm2, the results were within 2.0% of the measured values for all energies. Similarly, the values calculated using Monaco were within approximately 2.0% of the measured values for irradiation field sizes between 10 × 10 and 1.5 × 1.5 cm2 for all beam energies of interest. Thus, PSDs are effective validation tools for OPF calculations in TPS. A TPS modeled with the same source data has different minimum irradiation field sizes that can be calculated. These findings could aid in verification of equipment accuracy for treatment planning requiring highly accurate dose calculations and for third-party evaluation of OPF calculations for TPS.
|
キーワード | Plastic scintillator
Radiation therapy
Small irradiation field
Output factor
|
備考 | This version of the article has been accepted for publication, after peer review (when applicable) and is subject to Springer Nature’s AM terms of use, but is not the Version of Record and does not reflect post-acceptance improvements, or any corrections. The Version of Record is available online at: http://dx.doi.org/10.1007/s13246-024-01438-5
This fulltext file will be available in May 2025.
|
発行日 | 2024-05-16
|
出版物タイトル |
Physical and Engineering Sciences in Medicine
|
出版者 | Springer Science and Business Media LLC
|
ISSN | 2662-4729
|
資料タイプ |
学術雑誌論文
|
言語 |
英語
|
OAI-PMH Set |
岡山大学
|
著作権者 | © Australasian College of Physical Scientists and Engineers in Medicine 2024
|
論文のバージョン | author
|
PubMed ID | |
DOI | |
Web of Science KeyUT | |
関連URL | isVersionOf https://doi.org/10.1007/s13246-024-01438-5
|
Citation | Ando, Y., Okada, M., Matsumoto, N. et al. Evaluation of output factors of different radiotherapy planning systems using Exradin W2 plastic scintillator detector. Phys Eng Sci Med (2024). https://doi.org/10.1007/s13246-024-01438-5
|