抄録 | The present retrospective study assessed radiographs to determine socket wear in total hip arthroplasty
(THA) with 22-mm zirconia or COP (Cobalt-Chrome alloy rich in Cobalt and Phosphorous)
heads, and in cemented stems at more than 10 years after operation. Sockets of ultra high molecular
weight polyethylene were used in each of two THA groups (13 hips each) in a clinical trial in our hospital
between 1989 and 1990. Three observers carried out masked assessments of the radiographs.
Upon fi nal examination, there was no remarkable loosening in the zirconia or COP group, and no
case had required revision surgery as of 2005. There was a statistically signifi cant diff erence between
the 2 groups in average annual linear wear, at 0.093 mm/year and 0.046 mm/year in the zirconia and
COP groups, respectively. Volume wear and average annual volume wear were also signifi cantly
greater in the zirconia group despite its superior mechanical strength and toughness in vitro. Our
present fi ndings do not confi rm early expectations of lower wear in long-term results of 22-mm zirconia
femoral heads used in THA.
|
キーワード | total hip arthroplasty
zirconia head
COP head
polyethylene
wear
|
Amo Type | Original Article
|
出版物タイトル |
Acta Medica Okayama
|
発行日 | 2006-12
|
巻 | 60巻
|
号 | 6号
|
出版者 | Okayama University Medical School
|
開始ページ | 311
|
終了ページ | 318
|
ISSN | 0386-300X
|
NCID | AA00508441
|
資料タイプ |
学術雑誌論文
|
言語 |
英語
|
論文のバージョン | publisher
|
査読 |
有り
|
PubMed ID | |
Web of Science KeyUT |