An application of the load cell as a sensor in venous occlusion plethysmography, which is a well-established method for limb or digit blood-flow measurements is presented. In this method, the limb volume changes that follow venous occlusion are transferred into the volume change of a water pool. The hydrostatic pressure as well as the water surface level are measured and used for the calculation of the volume change. By using this method, the influence of water pressure on limb blood flow is avoided together with drift and leakage of the sensing element. The load cell has the advantage that it measures the weight of the displaced water volume, which simplifies the design principles of the plethysmography. The plethysmography is found to be sensitive, highly linear, and easy to handle. The results of using a manufactured water-displacement plethysmography to determine actual blood-flow measurements in forearms and fingers are presented

キーワード
biomedical measurement
blood
electric sensing devices
flow measurement
haemodynamics
level measurement
volume measurement
weighing
備考
Digital Object Identifier: 10.1109/19.65812
Published with permission from the copyright holder. This is the institute's copy, as published in Instrumentation and Measurement, IEEE Transactions on , Dec 1990, Volume 39, Issue 6, Pages 987-990.
Publisher URL:http://dx.doi.org/10.1109/19.65812
Copyright © 1990 IEEE. All rights reserved.
発行日
1990-06-23
出版物タイトル
Instrumentation and Measurement
39巻
6号
開始ページ
987
終了ページ
990
資料タイプ
学術雑誌論文
言語
英語
査読
有り
DOI
Submission Path
electrical_engineering/46