Background and purpose: The low tracking accuracy of lung stereotactic body radiotherapy (SBRT) risks reduced treatment efficacy. We used four-dimensional computed tomography (4DCT) images to determine the correlation between changes in fiducial marker positions and lung volume for lung tumors, and we evaluated the effectiveness of the combined use of these images in lung SBRT.
Materials and methods: Data of 30 patients who underwent fiducial marker placement were retrospectively analyzed. We calculated the motion amplitudes of the center of gravity coordinates of the lung tumor and fiducial markers in each phase and the ipsilateral, contralateral, and bilateral lung volumes using 4DCT. Moreover, we calculated the cross-correlation coefficient between the fiducial marker position and the lung volume changes waveform for the motion amplitude waveform of the lung tumor over three gating windows (all phases, ≤2 mm3, and ≤3 mm3).
Results: Compared with the lung volume, approximately 30 % of the fiducial markers demonstrated a low correlation with the lung tumor. In the ≤2 mm3 and ≤3 mm3 gating windows, the cross-correlation coefficients between the lung tumor and the optimal marker (r > 0.9: 83 % and 86 %) were significantly different for all fiducial markers (r > 0.9: 39 %, 53 %) and the ipsilateral (r > 0.9: 35 % and 40 %), contralateral (r > 0.9: 44 % and 41 %), and bilateral (r > 0.9: 39 % and 45 %) lung volumes.
Conclusions: Some of the fiducial markers showed a low correlation with the lung tumor. This study indicated that the combined use of lung volume monitoring can improve tracking accuracy.
キーワード
Fiducial marker
Respiratory gating method
Stereotactic body radiotherapy
Tumor tracking
Lung cancer
4DCT
発行日
2022-10-12
出版物タイトル
Physics and Imaging in Radiation Oncology
24巻
出版者
Elsevier B.V.
開始ページ
82
終了ページ
87
ISSN
24056316
資料タイプ
学術雑誌論文
言語
英語
OAI-PMH Set
岡山大学
著作権者
© 2022 The Author(s).
論文のバージョン
publisher
PubMed ID
DOI
Web of Science KeyUT
関連URL
isVersionOf https://doi.org/10.1016/j.phro.2022.10.001
ライセンス
http://creativecommons.org/licenses/by/4.0/