Akavipat, Phuping
Department of Anesthesiology, Prasat Neurological Institute
Hungsawanich, Nate
Department of Anesthesiology, Prasat Neurological Institute
Jansin, Rungruedee
Department of Anesthesiology, Prasat Neurological Institute
|
抄録 | In neurosurgery in particular, the recommended placement of electrodes for monitoring depth of anesthesia during surgery sometimes conflicts with the surgical site or patient positioning. Therefore, we proposed this study to evaluate the agreement and correlation of bispectral index values recorded from the usual frontal area and the alternate, post-auricular areas in neurosurgery patients. Thirty-four patients scheduled for neurosurgery under general anesthesia were included. Bispectral index (BIS) sensors were placed at both the frontal and post-auricular areas. The anesthesia given was clinically adjusted according to the frontal (standard) BIS reading. The BIS values and impedance were recorded;Pearsonʼs correlation and Bland-Altman plots were analyzed. The bias ± 2SD for the electrode placement before, during, and post-anesthesia were 0±23.32, 1.5±10.69, and 2.1±13.52, while the limits of agreement were -23.3 to 23.3, -12.2 to 9.2, and -17.7 to 13.5, respectively. The correlation coefficient between frontal- and post-auricular-area electrodes was 0.74 with a p-value
<0.001.The post-auricular placement of a BIS electrode is a practical alternative to frontal lobe placement. Nevertheless, proper electrode location is important to minimize error.
|
キーワード | Bland and Altman analysis
electroencephalogram
instrumentation
equipment
|
Amo Type | Original Article
|
出版物タイトル |
Acta Medica Okayama
|
発行日 | 2014-06
|
巻 | 68巻
|
号 | 3号
|
出版者 | Okayama University Medical School
|
開始ページ | 151
|
終了ページ | 155
|
ISSN | 0386-300X
|
NCID | AA00508441
|
資料タイプ |
学術雑誌論文
|
言語 |
英語
|
著作権者 | CopyrightⒸ 2014 by Okayama University Medical School
|
論文のバージョン | publisher
|
査読 |
有り
|
PubMed ID | |
Web of Science KeyUT |