著者
Akavipat, Phuping Department of Anesthesiology, Prasat Neurological Institute
Hungsawanich, Nate Department of Anesthesiology, Prasat Neurological Institute
Jansin, Rungruedee Department of Anesthesiology, Prasat Neurological Institute
抄録
In neurosurgery in particular, the recommended placement of electrodes for monitoring depth of anesthesia during surgery sometimes conflicts with the surgical site or patient positioning. Therefore, we proposed this study to evaluate the agreement and correlation of bispectral index values recorded from the usual frontal area and the alternate, post-auricular areas in neurosurgery patients. Thirty-four patients scheduled for neurosurgery under general anesthesia were included. Bispectral index (BIS) sensors were placed at both the frontal and post-auricular areas. The anesthesia given was clinically adjusted according to the frontal (standard) BIS reading. The BIS values and impedance were recorded;Pearsonʼs correlation and Bland-Altman plots were analyzed. The bias ± 2SD for the electrode placement before, during, and post-anesthesia were 0±23.32, 1.5±10.69, and 2.1±13.52, while the limits of agreement were -23.3 to 23.3, -12.2 to 9.2, and -17.7 to 13.5, respectively. The correlation coefficient between frontal- and post-auricular-area electrodes was 0.74 with a p-value <0.001.The post-auricular placement of a BIS electrode is a practical alternative to frontal lobe placement. Nevertheless, proper electrode location is important to minimize error.
キーワード
Bland and Altman analysis
electroencephalogram
instrumentation
equipment
Amo Type
Original Article
出版物タイトル
Acta Medica Okayama
発行日
2014-06
68巻
3号
出版者
Okayama University Medical School
開始ページ
151
終了ページ
155
ISSN
0386-300X
NCID
AA00508441
資料タイプ
学術雑誌論文
言語
英語
著作権者
CopyrightⒸ 2014 by Okayama University Medical School
論文のバージョン
publisher
査読
有り
PubMed ID
Web of Science KeyUT