ID | 53021 |
JaLCDOI | |
フルテキストURL | |
著者 |
Hirayama, Takahiro
Department of Emergency and Critical Care Medicine, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
Nagano, Osamu
Department of Disaster and Emergency Medicine, Kochi University
Shiba, Naoki
Department of Emergency and Critical Care Medicine, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
Yumoto, Tetsuya
Department of Emergency and Critical Care Medicine, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
ORCID
Kaken ID
publons
Sato, Keiji
Department of Emergency and Critical Care Medicine, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
Terado, Michihisa
Department of Emergency and Critical Care Medicine, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
publons
Ugawa, Toyomu
Department of Emergency and Critical Care Medicine, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
ORCID
publons
researchmap
Ichiba, Shingo
Department of Emergency and Critical Care Medicine, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences
Kaken ID
publons
researchmap
|
抄録 | In adult high-frequency oscillatory ventilation (HFOV), stroke volume (SV) and mean lung pressure (PLung) are important for lung protection. We measured the airway pressure at the Y-piece and the lung pressure during HFOV using a lung model and HFOV ventilators for adults (R100 and 3100B). The lung model was made of a 20-liter, airtight rigid plastic container (adiabatic compliance:
19.3ml/cmH2O) with or without a resistor (20cmH2O/l/sec). The ventilator settings were as follows:
mean airway pressure (MAP), 30cmH2O;frequency, 5-15Hz (every 1Hz);airway pressure amplitude (AMP), maximum;and % of inspiratory time (IT), 50% for R100, 33% or 50% for 3100B. The measurements were also performed with an AMP of 2/3 or 1/3 maximum at 5, 10 and 15Hz. The PLung and the measured MAP were not consistently identical to the setting MAP in either ventilator, and decreasing IT decreased the PLung in 3100B. In conclusion, we must pay attention to the possible discrepancy between the PLung and the setting MAP during adult HFOV.
|
キーワード | HFOV
mean lung pressure
mean airway pressure
|
Amo Type | Original Article
|
出版物タイトル |
Acta Medica Okayama
|
発行日 | 2014-12
|
巻 | 68巻
|
号 | 6号
|
出版者 | Okayama University Medical School
|
開始ページ | 323
|
終了ページ | 329
|
ISSN | 0386-300X
|
NCID | AA00508441
|
資料タイプ |
学術雑誌論文
|
言語 |
英語
|
著作権者 | CopyrightⒸ 2014 by Okayama University Medical School
|
論文のバージョン | publisher
|
査読 |
有り
|
PubMed ID | |
Web of Science KeyUT | |
関連URL | http://ousar.lib.okayama-u.ac.jp/metadata/53132
|