抄録 | Patient-specific quality assurance (QA) data must be migrated from outdated QA systems to new ones to produce objective results that can be understood by oncologists. We aimed to evaluate a method for obtaining a high correlation of dose distributions according to various gamma passing rates among two types of 2D detectors for the migration of patient-specific QA data of volumetric modulated arc therapy (VMAT). The patient-specific QA of 20 patients undergoing VMAT was measured in two different modes: standard single measurement (SM) mode and multiple merged measurements (MM) techniques using Ar-cCHECK (AC) and OCTAVIUS (OT). The correlation of the measured and calculated dose distributions was evaluated according to varying gamma passing rates (3%/3 mm, 2%/3 mm, 2%/2 mm, and 1%/1 mm). The gamma passing rates were analyzed using the Anderson-Darling normality test. Treatment plan dose dis-tributions were calculated by intentionally shifting the calculation isocenter position (x,y,z +/- 0.5, +/- 1.0, +/- 1.5, and +/- 2.0 mm). The highest correlation between the SM and MM was observed with a gamma passing rate of 1%/1 mm with AC (r = 0.866) and 3%/2 mm with OT (r = 0.916). However, SM and MM did not follow a normal distribution with a rate of 3%/2 mm in OT. The second-highest correlation was obtained with a rate of 2%/2 mm (r = 0.900). Among the two 2D detectors, the highest correlation be-tween the calculated and measured dose distributions was obtained for a gamma passing rate of 1%/1 mm using SM in AC and 2%/2 mm using MM in OT (r = 0.716). Adjusting the gamma passing rate and measurement mode of AC and OT resulted in higher correlations between measured and calculated dose distributions. The high correlation between different 2D detectors objectively indicated a potential mi-gration method. This enabled the sharing of more accurate patient-specific QA data from 2D detectors with different phantoms. A high correlation was observed between the two types of detectors in this study (r = 0.716); therefore, the proposed method should be useful for oncologists to share information regarding patient-specific QA for VMAT.
|
キーワード | Volumetric modulated arc therapy (VMAT)
Patient-specific quality assurance (QA)
2D detector
Gamma passing rate
|
備考 | © 2022 American Association of Medical Dosimetrists. This manuscript version is made available under the CC-BY-NC-ND 4.0 License. http://creativecommons.org/licenses/by-nc-nd/4.0/.
This is the accepted manuscript version. The formal published version is available at [https://doi.org/10.1016/j.meddos.2022.06.003] .
This full-text will be available in Oct. 2023.
|
発行日 | 2022-10-31
|
出版物タイトル |
Medical Dosimetry
|
巻 | 47巻
|
号 | 4号
|
出版者 | Elsevier BV
|
開始ページ | 329
|
終了ページ | 333
|
ISSN | 0958-3947
|
資料タイプ |
学術雑誌論文
|
言語 |
英語
|
OAI-PMH Set |
岡山大学
|
著作権者 | © 2022 American Association of Medical Dosimetrists.
|
論文のバージョン | author
|
PubMed ID | |
DOI | |
Web of Science KeyUT | |
関連URL | isVersionOf https://doi.org/10.1016/j.meddos.2022.06.003
|
ライセンス | http://creativecommons.org/licenses/by-nc-nd/4.0/.
|