抄録

A modeling technique of the minor loop using typical hysteresis loops is shown. The effect of the minor loop and eddy current in the pole piece of a permanent magnet type of MRI on the residual flux density of the probe coil is examined. It is illustrated that the change ΔB of residual flux density occurs due to the minor loop of the pole piece. It is also pointed out that the choice of time interval Δt is important in a nonlinear analysis considering the minor loop

キーワード
biomedical MRI
eddy currents
finite element analysis
magnetic flux
magnetic hysteresis
permanent magnets
備考
Digital Object Identifier: 10.1109/20.767404
Published with permission from the copyright holder. This is the institute's copy, as published in Magnetics, IEEE Transactions on , May 1999, Volume 35, Issue 3, Part 1, Pages 1893-1896 .
Publisher URL:http://dx.doi.org/10.1109/20.767404
Copyright © 1999 IEEE. All rights reserved.
発行日
1999-5
出版物タイトル
Magnetics
35巻
3号
開始ページ
1893
終了ページ
1896
資料タイプ
学術雑誌論文
言語
英語
査読
有り
DOI
Submission Path
electricity_and_magnetism/19