ID | 30281 |
フルテキストURL | |
著者 |
Suenaga, Ryousuke
Okayama University
Miyata, Koji
Shin-Etsu Chemical Corporation Limited
Ohashi, Ken
Shin-Etsu Chemical Coporation Limited
|
抄録 | The minor loops of B and H of steel due to pulse excitation and eddy currents induced in steel affect the magnetic characteristics of a permanent-magnet-type MRI. In this paper, the magnetic properties of a permanent magnet assembly is examined by using the finite-element method taking into account minor loop. The distribution of residual magnetization in the yoke is illustrated, and the effect of residual magnetization on the behavior of residual flux density is examined. It is shown that the behavior of B and H in minor loops is affected by the eddy currents in the yoke and pole piece. |
キーワード | biomedical MRI
eddy currents
finite element analysis
magnetic hysteresis
magnetisation
permanent magnets
steel
|
備考 | Digital Object Identifier: 10.1109/TMAG.2003.810510
Published with permission from the copyright holder. This is the institute's copy, as published in Magnetics, IEEE Transactions on, May 2003, Volume 39, Issue 3, Part 1, Pages 1337-1340. Publisher URL:http://dx.doi.org/10.1109/TMAG.2003.810510 Copyright © 2003 IEEE. All rights reserved. |
発行日 | 2003-5
|
出版物タイトル |
Magnetics
|
巻 | 39巻
|
号 | 3号
|
開始ページ | 1337
|
終了ページ | 1340
|
資料タイプ |
学術雑誌論文
|
言語 |
英語
|
査読 |
有り
|
DOI | |
Submission Path | electricity_and_magnetism/44
|