抄録
This study presents a new method for measuring the Seebeck coefficient under high pressure in a multi-anvil apparatus. The application of a dual-heating system enables precise control of the temperature difference between both ends of the sample in a high-pressure environment. Two pairs of W–Re thermocouples were employed at both ends of the sample to monitor and control the temperature difference, and independent probes were arranged to monitor the electromotive force (emf) produced by temperature oscillation at a given target temperature. The temperature difference was controlled within 1 K during the resistivity measurements to eliminate the influence of the emf owing to a sample temperature gradient. The Seebeck measurement was successfully measured from room temperature to 1400 K and was obtained by averaging the two measured values with opposite thermal gradient directions (∼20 K). Thermoelectric properties were measured on disk-shaped p-type Si wafers with two different carrier concentrations as a reference for high Seebeck coefficients. This method is effective to determine the thermoelectric power of materials under pressure.
発行日
2020-03-19
出版物タイトル
Review of Scientific Instruments
91巻
3号
開始ページ
035115
ISSN
00346748
NCID
AA00817730
資料タイプ
学術雑誌論文
言語
英語
OAI-PMH Set
岡山大学
論文のバージョン
author
PubMed ID
DOI
Web of Science KeyUT
関連URL
isVersionOf https://doi.org/10.1063/1.5143525
助成機関名
文部科学省
助成番号
15H05827
17H01155
17K06847