抄録
During radiation therapy, incident radiation oblique to the skin surface is high and may cause severe skin damage. Understanding the dose of radiation absorbed by the skin is important for predicting skin damage due to radiation. In this study, we used a high-energy (4 MV) X-ray system and an optically stimulated luminescence dosimeter (OSLD) that was developed for personal exposure dosimetry. We determined the dose variation and angular dependence, which are the characteristics of a small OSLD required to derive the calculation formula for the oblique surface dose. The dose variation was determined using the coefficient of variation. The maximum coefficient of variation for 66 small-field OSLDs was 1.71%. The angular dependence, obtained from the dose ratio of the dosimeter in the vertical direction, had a maximum value of 1.37. We derived a new equation in which the oblique surface dose can be calculated within the error range of −7.7-5.1%.
キーワード
optically stimulated luminescent dosimeter
radiotherapy
oblique surface dose
high-energy X-ray therapy
angular dependence
Amo Type
Original Article
出版物タイトル
Acta Medica Okayama
発行日
2020-10
74巻
5号
出版者
Okayama University Medical School
開始ページ
415
終了ページ
422
ISSN
0386-300X
NCID
AA00508441
資料タイプ
学術雑誌論文
言語
英語
著作権者
CopyrightⒸ 2020 by Okayama University Medical School
論文のバージョン
publisher
査読
有り
PubMed ID
Web of Science KeyUT
NAID