抄録 | A method for the on-site estimation of arsenic (As) contamination in soil by solid-phase extraction with a resin based on molecular recognition technology (SPE-MRT) and anodic stripping voltammetry (ASV) using a gold film microelectrode was developed. Optimized pretreatments were carried out as follows: 1 g of soil sample was taken and put into polypropylene test tube containing 33 mL 1 M hydrochloric acid solution, which was heated at 70 degree Celsius for 5 min using graphite block hot plate. As in the sample was extracted by shaking for 1 min using an automatic mini-shaker system. The extract was finally filtered through a 0.45 μm disk PTFE membrane filter. The extracted arsenic was electrodeposited on the gold film carbon microelectrode at −400 mV vs. Ag/AgCl for 1 min and the deposits were stripped at a rate of 5 mV/s to 580 mV vs. Ag/AgCl by using AnaLig® TE-3, Cu in soil extract was removed with 1 M hydrochloric acid for the determination of As. The method can eliminate Cu interference. The detection limit was 0.5 μg/L in soil extract. The results of voltammograms for As in the extracts from real soil samples were the same level as these obtained by a cross-check method using inductively coupled plasma mass spectroscopy (ICP-MS).
|
キーワード | solid phase extraction
anodic stripping voltammetry
soil extract
arsenic
|
発行日 | 2008-01-29
|
出版物タイトル |
分析化学
|
出版物タイトル(別表記) | Bunseki Kagaku
|
巻 | 56巻
|
号 | 12号
|
出版者 | 日本分析化学会
|
出版者(別表記) | The Japan Society for Analytical Chemistry
|
開始ページ | 1165
|
終了ページ | 1169
|
ISSN | 0525-1931
|
NCID | AN00222633
|
資料タイプ |
学術雑誌論文
|
オフィシャル URL | http://dx.doi.org/10.2116/bunsekikagaku.56.1165
|
言語 |
日本語
|
著作権者 | 日本分析化学会
|
論文のバージョン | publisher
|
査読 |
有り
|
DOI |