Memoirs of the Faculty of Engineering, Okayama University volume25 issue1
1990-12-14 発行
Is proposed a tunable diode laser absorption spectrometer system which employs a pulsed current to drive the diode laser still implementing a second-harmonic detection methods enhancing a signal to noise up to 104 times that the case without it. This system affords the lead-salt diode laser a higher operating temperature which allows a more compact deep cooling system. A principle is based on employing a gated integrator between the preamplifier and the lock-in amplifier. Investigations are made on the optimum selection of the gating aperture time interval as determined by response time of the infrared detector and the pulse width of the laser driving current.