ID | 21163 |
JaLCDOI | |
Sort Key | 8
|
タイトル(別表記) | Determination of lithium in serum by emission spectrophotometry using "Spectraspan"
|
フルテキストURL | |
著者 |
御舩 政明
岡山大学医学部附属病院三朝分院中央検査部
|
抄録 | Lithium in serum was determined by a plasma emissionspectrophotometer "Spectraspan" (low power d.c. plasma are, operating on argon). The opitimum condition for ashing of serum by low temperature was studied. The influences of hydrochloric acid and several metallic ions (Na, K, Mg, Ca) on emission intensity were examined. There was a linear relationship between emission intensity and lithium content from 0 to 10μg/ml. Lithum in serum was easily determined by standard addition method. Precision, coefficient of variance and recovery of known amount of lithium added to the sample for 12 replicate analyses were 0.093μg/ml, 1.85%, 99-104%, respectively. Significantly positive correlationship was observed between emission spectrophotometry using spectraspan and atomic absoption spectrophotometry.
|
備考 | 原著論文 (Original Papers)
|
出版物タイトル |
岡山大学温泉研究所報告
|
発行日 | 1980-03-25
|
巻 | 50巻
|
出版者 | 岡山大学温泉研究所
|
出版者(別表記) | Institute for Thermal Spring Research, Okayama University
|
開始ページ | 43
|
終了ページ | 48
|
ISSN | 0369-7142
|
NCID | AN00032853
|
資料タイプ |
紀要論文
|
OAI-PMH Set |
岡山大学
|
言語 |
日本語
|
論文のバージョン | publisher
|
NAID | |
Eprints Journal Name | mmc
|