Abstract | A highly sensitive fluorescence-quenching method for the determination of phosphorus based on the formation of an ion associate between molybdophosphate and Rhodamine B (RB) was developed. A simple flow-injection system coupled with a fluorescence detector was used to measure the fluorescence intensity at 560 nm and 580 nm as an excitation and an emission wavelength, respectively. The calibration graph for phosphorus showed a good linearity in the range of (0-1) x 10(-7) M (1 M = 1 mol L-1), and a detection limit of I x 10(-9) M (SIN = 3). The proposed method was successfully applied to the determination of ultratrace amounts of phosphorus in ultrapurified and purified water samples, and to the determination of trace amounts of phosphorus in commercially-available hydrogen peroxide solutions with satisfactory results. |
Note | Published with permission from the copyright holder.
This is published in Analytical Sciences, 2005 Vol.21 Issue.3 pp.263-268 Publisher URL: http://dx.doi.org/10.2116/analsci.21.263 Direct access to Thomson Web of Science record (C)2005 by The Japan Society for Analytical Chemistry |
Published Date | 2005-03
|
Publication Title |
Analytical Sciences
|
Volume | volume21
|
Issue | issue3
|
Publisher | The Japan Society for Analytical Chemistry
|
Publisher Alternative | 日本分析化学会
|
Start Page | 263
|
End Page | 268
|
ISSN | 0910-6340
|
NCID | AA10500785
|
Content Type |
Journal Article
|
language |
English
|
Copyright Holders | The Japan Society for Analytical Chemistry
|
File Version | publisher
|
Refereed |
True
|
DOI | |
Web of Science KeyUT | |
Submission Path | analytical_chemistry/45
|