このエントリーをはてなブックマークに追加
ID 20495
Eprint ID
20495
FullText URL
Title Alternative
Relationship between X-ray Exposure and Patient Dose on the Radiological Diagnosis
Author
Nakagiri, Yoshitada
Sibuya, Koichi
Maruyama, Toshinori Kaken ID publons researchmap
Sugita, Katsuhiko
Sonoyama, Masayo
Morioka, Yasuki
Inamura, Keiji
Tahara, Seiji
Uno, Hirofumi
Kadohisa, Shigefumi
Inoue, Tatsuya
Hiraki, Yoshio
Abstract
Patient dose is important consideration in the radiological examination and our environment regarding radiation. Many studies have been published about patient dose, but those data were classified by each organ or tissue. Actuarially, patient dose should be checked by each examined part of patient and each exposure equipment. In this paper, we measured absorbed dose at the depth of 0-200mm with the Mix-DP phantom. The phantom is made by tissue equivalent meterial and is designed to similitude abdominal part. Percentage Depth Dose (PDD) was calculated from these doses. Three single-phase generators and three three-phases generators were used in this measurement. These measurements were analyzed by each equipment, and consequently the clear difference of PDD between the exposure equipments was not found. As the result, we can estimate patient dose at a random depth by using PDD. Furthermore, we can easily know patient dose from the tube-voltage and current time product by the calculation including PDD.These data are very useful to manager patient dose on radiological diagnosis.
Keywords
医用放射線被曝 (patient dose)
X線撮影条件 (X-ray exposure)
線量測定 (Radiation dosimetry)
Note
論文
Published Date
1995-12-20
Publication Title
環境制御
Publication Title Alternative
Environmental Research and Control
Volume
volume17
Publisher
岡山大学環境管理センター
Publisher Alternative
Okayama University Administration Center for Environmental Science and Technology
Start Page
63
End Page
70
ISSN
0917-1533
NCID
AN1024789X
Content Type
Departmental Bulletin Paper
language
Japanese
File Version
publisher
Refereed
True
Eprints Journal Name
erc