このエントリーをはてなブックマークに追加
ID 48688
JaLCDOI
FullText URL
Author
Shiba, Naoki
Nagano, Osamu
Hirayama, Takahiro
Ujike, Yoshihito Kaken ID publons
Abstract
In adult high-frequency oscillatory ventilation (HFOV) with an R100 artificial ventilator, exhaled gas from patientʼs lung may warm the temperature probe and thereby disturb the humidification of base flow (BF) gas. We measured the humidity of BF gas during HFOV with frequencies of 6, 8 and 10Hz, maximum stroke volumes (SV) of 285, 205, and 160ml at the respective frequencies, and, BFs of 20, 30, 40l/min using an original lung model. The R100 device was equipped with a heated humidifier, HummaxⅡ, consisting of a porous hollow fiber in circuit. A 50-cm length of circuit was added between temperature probe (located at 50cm proximal from Y-piece) and the hollow fiber. The lung model was made of a plastic container and a circuit equipped with another HummaxⅡ. The lung model temperature was controlled at 37℃. The HummaxⅡ of the R100 was inactivated in study-1 and was set at 35℃ or 37℃ in study-2. The humidity was measured at the distal end of the added circuit in study-1 and at the proximal end in study-2. In study-1, humidity was detected at 6Hz (SV 285ml) and BF 20l/min, indicating the direct reach of the exhaled gas from the lung model to the temperature probe. In study-2 the absolute humidity of the BF gas decreased by increasing SV and by increasing BF and it was low with setting of 35℃. In this study setting, increasing the SV induced significant reduction of humidification of the BF gas during HFOV with R100.
Keywords
HFOV
humidification
Amo Type
Original Article
Publication Title
Acta Medica Okayama
Published Date
2012-08
Volume
volume66
Issue
issue4
Publisher
Okayama University Medical School
Start Page
335
End Page
341
ISSN
0386-300X
NCID
AA00508441
Content Type
Journal Article
language
English
Copyright Holders
CopyrightⒸ 2012 by Okayama University Medical School
File Version
publisher
Refereed
True
PubMed ID
Web of Science KeyUT