Automatic measurement of oxygen isotope ratio of water samples
FullText URL
Author
Chiba, Hitoshi Kaken ID researchmap
Sakai, Hitoshi
Yasutake, Masatoshi
Abstract
The automatic sample preparation system for oxygen isotope analysis of natural water samples was constructed. The system is essentially a modification of that originally designed by W. Dansgaard in the University of Copenhagen. Sixty water samples of 5 ml each are automatically equilibrated with CO(2) of 30 ml NTP each within 4.5 hours. The equilibrated CO(2) gases are successively measured for their oxygen isotopic ratios by an automatic mas-sspectrometer, VG903. The time required for the measurement of the 60 CO(2) samples is about 15 hours; thus approximately 20 hours are required to complete the isotopic measurements of 60 water samples. The accuracy of the oxygen isotopic analyses is about 0.1%. The result of oxygen isotopic analyses by the automatic preparation system agrees with that of the conventional method within about +0.1%. The results indicate that the automatic preparation system is applicable for the oxygen isotope analysis of natural water samples.